Measuring the LD Current
Overview
Section titled “Overview”The measurement of the laser diode current can be done by measuring the reference voltage at the points above the shunt resistor while the laser diode is connected.
This provides a non-invasive method to monitor the actual laser current during operation.
Measurement Points Location
Section titled “Measurement Points Location”The measurement points are located above the shunt resistor on the driver board.
┌─────────────┐ │ Measure │ │ Points │ └──────┬──────┘ │ ┌──────┴──────┐ │ Shunt │ │ Resistor │ └─────────────┘ │ To LaserMeasurement Procedure
Section titled “Measurement Procedure”Required Equipment
Section titled “Required Equipment”- Digital multimeter (DMM)
- Voltage measurement capability
- Minimum 3½ digit resolution
- DC voltage range selection
Step-by-Step Measurement
Section titled “Step-by-Step Measurement”-
Ensure laser diode is connected to the driver
-
Power up the driver and enable laser output
- Close interlock
- Set start signal (if required)
- Apply modulation voltage
-
Set multimeter to DC voltage mode
- Select appropriate range (typically 200mV or 2V)
- Ensure DC coupling
-
Connect multimeter probes to measurement points
- Red probe: Upper measurement point (positive)
- Black probe: Lower measurement point (negative/ground)
-
Read voltage displayed on multimeter
-
Calculate current using the conversion ratio for your model
Voltage-to-Current Conversion
Section titled “Voltage-to-Current Conversion”The conversion ratio is model-dependent based on the shunt resistor value.
Conversion Ratios
Section titled “Conversion Ratios”| Model | Conversion Ratio | Calculation |
|---|---|---|
| iLD500 | 1.0 V/A (1:1) | I(A) = V(V) × 1 |
| iLD2500 | 0.2 V/A (1:5) | I(A) = V(V) × 5 |
| iLD4500 | 0.1 V/A (1:10) | I(A) = V(V) × 10 |
Formula by Model
Section titled “Formula by Model”iLD500 (10-500mA)
Section titled “iLD500 (10-500mA)”I_LD (A) = V_measured (V) × 1I_LD (mA) = V_measured (V) × 1000iLD2500 (70-2500mA)
Section titled “iLD2500 (70-2500mA)”I_LD (A) = V_measured (V) × 5I_LD (mA) = V_measured (V) × 5000iLD4500 (140-4500mA)
Section titled “iLD4500 (140-4500mA)”I_LD (A) = V_measured (V) × 10I_LD (mA) = V_measured (V) × 10000Worked Examples
Section titled “Worked Examples”Example 1: iLD500
Section titled “Example 1: iLD500”Measured voltage: 0.04 V
Calculation:
I_LD = 0.04 V × 1 = 0.04 A = 40 mAResult: Laser current is 40 mA
Example 2: iLD2500
Section titled “Example 2: iLD2500”Measured voltage: 0.04 V
Calculation:
I_LD = 0.04 V × 5 = 0.20 A = 200 mAResult: Laser current is 200 mA
Example 3: iLD4500
Section titled “Example 3: iLD4500”Measured voltage: 0.04 V
Calculation:
I_LD = 0.04 V × 10 = 0.40 A = 400 mAResult: Laser current is 400 mA
Example 4: High Current (iLD4500)
Section titled “Example 4: High Current (iLD4500)”Measured voltage: 0.35 V
Calculation:
I_LD = 0.35 V × 10 = 3.50 A = 3500 mAResult: Laser current is 3500 mA (3.5 A)
Measurement Range by Model
Section titled “Measurement Range by Model”Expected Voltage Ranges
Section titled “Expected Voltage Ranges”| Model | Current Range | Voltage Range |
|---|---|---|
| iLD500 | 10 - 500 mA | 0.01 - 0.5 V |
| iLD2500 | 70 - 2500 mA | 0.014 - 0.5 V |
| iLD4500 | 140 - 4500 mA | 0.014 - 0.45 V |
Quick Reference Table
Section titled “Quick Reference Table”iLD500 (1:1 ratio)
Section titled “iLD500 (1:1 ratio)”| Voltage (V) | Current (mA) |
|---|---|
| 0.010 | 10 |
| 0.050 | 50 |
| 0.100 | 100 |
| 0.200 | 200 |
| 0.300 | 300 |
| 0.400 | 400 |
| 0.500 | 500 |
iLD2500 (1:5 ratio)
Section titled “iLD2500 (1:5 ratio)”| Voltage (V) | Current (mA) |
|---|---|
| 0.014 | 70 |
| 0.100 | 500 |
| 0.200 | 1000 |
| 0.300 | 1500 |
| 0.400 | 2000 |
| 0.500 | 2500 |
iLD4500 (1:10 ratio)
Section titled “iLD4500 (1:10 ratio)”| Voltage (V) | Current (mA) |
|---|---|
| 0.014 | 140 |
| 0.100 | 1000 |
| 0.200 | 2000 |
| 0.300 | 3000 |
| 0.400 | 4000 |
| 0.450 | 4500 |
Measurement Best Practices
Section titled “Measurement Best Practices”Accuracy Considerations
Section titled “Accuracy Considerations”Factors affecting measurement accuracy:
- Multimeter accuracy (typically ±0.5% to ±2%)
- Contact resistance at measurement points
- Shunt resistor tolerance
- Temperature effects on shunt resistance
Expected accuracy: ±2-5% typical
Improving Accuracy
Section titled “Improving Accuracy”- ✅ Use quality digital multimeter
- ✅ Ensure good probe contact
- ✅ Allow system to stabilize thermally
- ✅ Take multiple readings and average
- ✅ Calibrate against known current source (if available)
Measurement Tips
Section titled “Measurement Tips”Alternative Measurement Methods
Section titled “Alternative Measurement Methods”External Current Meter
Section titled “External Current Meter”You can also measure current by inserting a current meter in series with the laser diode:
Oscilloscope Measurement
Section titled “Oscilloscope Measurement”For dynamic current measurements:
- Connect oscilloscope probe to measurement points
- Set to DC coupling
- Use appropriate voltage range
- Apply conversion ratio to waveform
Advantages:
- View current waveform in real-time
- Measure modulation depth
- Analyze rise/fall times
- Detect current ripple
Troubleshooting Measurements
Section titled “Troubleshooting Measurements”Zero or Very Low Voltage
Section titled “Zero or Very Low Voltage”Possible Causes:
- Laser not enabled
- No modulation voltage
- Poor probe contact
- Measurement points not connected
Solutions:
- Verify laser is enabled (Laser LED ON)
- Check modulation voltage >0.3V
- Clean measurement points
- Ensure good probe contact
Voltage Higher Than Expected
Section titled “Voltage Higher Than Expected”Possible Causes:
- Current set too high
- Wrong conversion ratio used
- Meter on wrong range
Solutions:
- Verify GAIN setting
- Check driver model and use correct ratio
- Select appropriate meter range
Unstable Reading
Section titled “Unstable Reading”Possible Causes:
- Modulation signal noise
- Poor probe contact
- Thermal effects
- Power supply ripple
Solutions:
- Stabilize modulation source
- Improve probe contact
- Allow thermal stabilization
- Check power supply quality
Reading Doesn’t Match Expected
Section titled “Reading Doesn’t Match Expected”Possible Causes:
- Wrong model assumed
- BIAS/GAIN not set correctly
- Calculation error
Solutions:
- Verify driver model number
- Re-set BIAS and GAIN
- Double-check conversion ratio
Safety Precautions
Section titled “Safety Precautions”During Measurement
Section titled “During Measurement”- ✅ Ensure laser safety precautions (goggles, enclosure)
- ✅ Avoid touching high-voltage points
- ✅ Keep hands clear of laser beam path
- ✅ Monitor laser temperature
- ✅ Watch for overheating indicators
Continuous Monitoring
Section titled “Continuous Monitoring”For applications requiring continuous current monitoring:
Data Logger Connection
Section titled “Data Logger Connection”Connect a data acquisition system to measurement points:
- Use high-impedance input (>1MΩ)
- Sample at appropriate rate
- Apply conversion ratio in software
- Log to file for analysis
Panel Meter Installation
Section titled “Panel Meter Installation”Install analog or digital panel meter:
- Connect across measurement points
- Calibrate using conversion ratio
- Provides visual indication
- Can add alarm outputs
Documentation
Section titled “Documentation”Recording Measurements
Section titled “Recording Measurements”Document current measurements including:
- Date and time
- Modulation voltage
- Measured voltage
- Calculated current
- Operating conditions (temperature, etc.)
- Any observations
This creates a valuable operating history for:
- Troubleshooting
- Trend analysis
- Laser aging assessment
- Performance verification